Description
Breas Clearway 2 Technical Specifications
|
Settings / Performance |
|
|
Ventilation Modes |
Manual, Basic Auto, Program Mode, IPPB, NIV |
|
Device Modes |
Home/Clinical |
|
Settings Profiles |
Four profiles |
|
Insufflation Pressure |
3 cmH2O–70 cmH2O |
|
Exsufflation Pressure |
-3 cmH2O– -70 cmH2O |
|
Insufflation Time |
0.5–5 s |
|
Exsufflation Time |
0.5–5 s |
|
Pause Time |
0–9 s (Auto Mode 0.5–9 s) |
|
Flow Options (MI-E) |
1–10 (10–100% of Ti max 4 s) |
|
Flow (IPPB) |
20 L/ min–120 L/ min |
|
Inspiratory Trigger |
Off, 1–10 |
|
Oscillations Frequency |
Insufflations and/ or Exsufflations, Off−1−20 Hz |
|
Oscillations Amplitude |
1 cmH2O–10 cmH2O |
|
Monitoring |
|
|
Displayed Data |
Cough Peak Flow, Insufflation Volume, Pressure, Flow, SpO2, Heart Rate Sessions Used, Average Session Duration |
|
Waveforms |
Pressure |
|
Trends and Statistical Data via PC Software |
Cough Peak Flow, Insufflation Volume, Use, SpO2, Heart Rate |
|
Power Supplies |
|
|
Mains Supply |
100–240 V AC |
|
Internal Battery |
>30 min in MI-E Modes (based on settings of +40 / –45 cmH2O, Ti/ Te / P 2 s) |
|
Alarms |
|
|
Alarms |
Circuit Disconnection, Running on Battery Power, External Battery Fault, High Pressure Alarm, Internal Over Temperature Alarm, Low Battery, Low Memory Battery, Machine Fault, Low SpO2 Reading, High SO2, SpO2 Sensor Disconnect, High Heart Rate, Low Heart Rate, Power Fail |
|
Dimensions |
|
|
W x H x D |
285 x 285 x 195 mm |
|
Weight |
3.75 kg |
|
Noise Level (at 10 cmH2O constant pressure) |
Sound level at 70 cmH2O in Manual Mode less than 74 dB(A) measured at 1 m |
|
Connectivity |
|
|
Built-in Connectivity |
USB, SD card |
|
Plug-and-Play Accessories |
SpO2, Hand Held Remote Start/Stop and Manual Switch |
|
Other |
|
|
Circuits |
Disposable circuit with bacterial filter no leak port for MI-E Modes and disposable circuit with bacterial filter and leak port for IPPB and NIV Modes |